Dew Point Set DS 400 for Monitoring of Medical Compressed Air

Compared to DIN EN 737-3, the “Aer medicalis” monograph included in the appendix to the European Pharmacopoeia presents substantially modified purity characteristics. Particularly noteworthy is the significant reduction of the dew point temperature to -46°C atmospheric (equivalent to 67 ppm V/V), which presents considerable challenges. The value of 67 ppm V/V for water is equivalent to the dew point at a pressure of 200 bars and a temperature of 11ºC. The need to limit the water content to the condensation point is crucial, as water condensation could lead to corrosion damage and promote microbial growth.

TECHNICAL DATA DS 400

 

Dimensions: 118 x 115 x 98 mm

IP 54 (wall housing)

92 x 92 x 75 mm

(panel mounting)

Inputs: 2 digital inputs for FA 510

2 analogue inputs for pressure sensors

Interface: USB-interface
Power supply: 100…240 VAC, 50-60 Hz
Accuracy: See technical data FA 510
Alarm outputs: 2 relays, (pot.-free)
Data logger:

 

100 million measured values start/stop time, measuring rate freely adjustable

 