Key features include:
- A 3-in-1 sensor to measure absolute pressure, temperature, and humidity in the room
- Interfaces for Modbus-RTU, Ethernet, or M-Bus
- An alarm relay with limit value adjustable via buttons (max 60 VDC, 0.5 A)
- Optional: 2 x 4…20 mA analogue outputs and 2 x alarm relays, suitable for dew point and temperature, for instance.
TECHNICAL DATA of IAC 500
| Temperature
Measuring range: |
-20…+60°C |
| Accuracy: | ± 1.0 K (0…60°C)
± 1.25 K (-20…0°C) |
| Rel. humidity
Measuring range: |
5…95% rF |
| Accuracy: | ± 3% |
| Absolute pressure
measuring range: |
300…1100 hPa (abs) |
| Accuracy: | ±1.7 hPa |
| Voltage supply: | 24 VDC (18…30 VDC via
SELV supply) or Power over Ethernet (IEEE802.3af: Class 2 (3.84W – 6.49W) |
| Protection class of
housing: |
IP 65 |
| Operating temperature: | -20…+60°C |
| Ambient temperature: | -20…+60°C |
| Storage temperature: | -40…+80°C |
| Readable via
Modbus: |
Absolute pressure [hPa, mbar, bar, psi,…],
dew point [°Ctd, °Ftd], temperature [°C, °F], rel. humidity [ %RH], abs. humidity [g/m³]. |