TECHNICAL DATA
Measuring range: | -80…20 °Ctd, -20…50 °Ctd |
Accuracy: | ± 1 °C at 50…-20 °Ctd ± 2 °C at -20…-50 °Ctd ± 3 °C at -50…-80 °Ctd |
Pressure range: | -1…50 bar Special version up to 500 bar |
Power supply: | 24 VDC (10…36 VDC) |
Protection class: | IP 66 |
EMC: | In acc. with DIN EN 61326-1 |
Operating temperature: | -20…70 °C |
Connection: | M12, 5-pin |
PC connection: | Modbus-RTU interface (RS485) |
Analogue output: | 4…20 mA = -80…20 °Ctd 4…20 mA = -20…50 °Ctd FA 510: 4…20 mA (3-wire) FA 515: 4…20 mA (2-wire) |
Burden for analogue output: | < 500 Ω |
Screw-in thread: | G 1/2″ Stainless steel Optional: UNF 5/8″, NPT 1/2″ |
Dimensions: | Ø 30 mm, length approx. 130 mm |
Via service software: Choose units |
% RH, °Ctd, g/m³, mg/m³, ppm V/V |
Scaling | 4…20 mA change |